The Smiths Medical CADD Legacy Plus 6500 (#CADD-6500) includes the pump, a single use 50/100 ml pump pouch, two AA alkaline batteries, non-sterile demo cassette, the operator’s manual, a patient guide, and a carrying case. Specially designed for continuous and intermittent therapies, this state-of-the-art pump delivers medication for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, or subarachnoid space infusion.

Category: Brand:
No Sales Tax
1 year warranty on equipment orders
Commercial financing available for orders over $2,000

Cadd Legacy Plus 6500

Refurbished with 1 year warranty

The CADD®-Legacy PLUS Model 6500  pump is the new member of the CADD®-Legacy pump family of therapy-specific delivery systems.

The CADD® line of pumps sets the standard for continuous and intermittent ambulatory infusion in research.
CADD® pumps have an unparalleled track record of accuracy and durability dating back to the early 1980’s.
These pumps withstand the rigors of primate and canine ambulatory like no other!

  • Infusion rates up to 125 ml/hr and dose volumes up to 1,000 ml can accommodate high-rate, high-volume therapies
  • Longer delayed start time and expanded dosing parameters provide added flexibility in meeting complex therapy requirements
  • Internal clock designed to keep dosing on schedule

Easy to read display

  • Cassette detection.
  • Occlusion detection.
  • Air-in-line detection.
  • On/off key.
  • Event memory.
  • Reliability.
  • Programming flexibility.

Upstream and downstream occlusion sensors

ON/OFF key allows pump to go in low energy mode to save battery energy
Event memory: last 1,000 events can be downloaded to a PC with CADD® Diplomat Software
Infusion rates: up to 125 mL/hr
Dose volumes to 1,000 mL
Internal clock for dosing schedule
Re-certified through a multi-step process involving biomedical checks and computer simulations
Fully refurbished and patient ready.
We offer these pumps with a 3 month warranty option a 6 month warranty option as well as a 1 year warranty option,please contact us for a competitive quote.

Description

 

Pump Resolution:
0.050 ml per pump stroke nominal

 

Dimensions:
1.6″ D x 3.8″ W x 4.4″ H excluding cassette

 

Weight:
13.8 oz including 2 AA batteries and empty 100 ml reservoir

 

Power Source:
2 AA alkaline batteries or AC Adapter

 

Maximum Infusion Pressure:
40.0 psi

 

System Delivery Accuracy:
±6% nominal

 

Maximum Time to Occlusion Alarm:
2.0 hours

 

Delivery Rate During Priming:
Approximately 180 ml/hr
Continuous Delivery Mode Specifications

 

Reservoir Volume:
1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments

 

Continuous Rate:
0.1 – 125 ml/hr; programmable in 0.1 ml/hr increments

 

Given:
0-99999.95 in 0.05 ml increments
Intermittent Delivery Mode Specifications

 

Reservoir Volume:
1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments

 

Dose Volume:
0.1 – 1000 ml programmable in 0.1 ml increments

 

Dose Duration:
1 min to 24 hrs in the following increments
– 1 minute for values between 1 min and 10 min
– 5 minutes for values above 10 min

 

Dose Cycle:
10 min to 96 hrs in 5 min increments

 

KVO Rate:
0 to 125 ml/hr in 0.1 ml/hr increments

 

Dose Starts In:
Immediate or 1 min to 96 hrs in the following increments
– 1 min for values between 0 min and 10 min
– 5 min for values between 10 min and 96 hrs

 

Given:
0-99999.95 in 0.05 ml increments

 

Dose Remaining:
1 minute increments

 

Pump Alarms:
Low battery power
Depleted battery power
Battery dislodged
Pump stopped
Pump fault
Low reservoir volume
High delivery pressure
Air in line
Disposable not attached when run attempted
Motor locked
Upstream occlusion
Reservoir volume empty
Program incomplete
Key stuck
Disposable detached

CADD-Legacy-Plus-6500-Ops-Manual